N/A in certain parameter in “Results”
Possible Cause:
Certain parameters are rejected due to improper sample, improper card, or defects on the card.
Action Required:
Use a new Q-Card, repeat the measurement.
N/A in certain parameter in “Results”
Possible Cause:
Certain parameters are rejected due to improper sample, improper card, or defects on the card.
Action Required:
Use a new Q-Card, repeat the measurement.
Abnormal noise from inside the device
Possible Cause:
Foreign matter is present inside the device.
Action Required:
Please turn off the power and unplug the device. Check the Q-Card compartment and heat dissipation holes, then remove any foreign matter.
N/A in certain parameter in “Results” again
Possible Cause:
Improper sample for iMOST.
Action Required:
Sample this parameter is improper to be tested in iMOST system.
Warning: Not enough blood
Possible Cause:
Insufficient blood sample in the Q-Card.
Action:
Ensure that an adequate blood sample is placed in the Q-Card.
Warning: Not fully insert Q-Card
Possible Cause:
The Q-Card is not fully inserted into the slot.
Action:
Ensure that the Q-Card is fully inserted until a physical stop is reached.
Warning: wrong Q-Cards
Possible Cause:
No Q-Card inserted, incorrect Q-Card inserted, or Q-Card inserted incorrectly.
Action:
Ensure the correct Q-Card is properly inserted.